Deliberations When Using Interlock ECG Patient Cables with Medical Machines

Disclaimer: This is a user generated content submitted by a member of the WriteUpCafe Community. The views and writings here reflect that of the author and not of WriteUpCafe. If you have any complaints regarding this post kindly report it to us.

Some hospital clients may wish to interface the defibrillator/monitor with other medical machines. Usually denoted as “slaving”, the patient’s ECG signal is cabled from the analog production of one machine to the patient effort of another, for remote scrutinizing. 

Appraisal: What is Coordinated Cardioversion?

 With the SYNC style designated on the defibrillator/monitor, each QRS intricate the patient fabricates is recognized and marked. The defibrillator only permits the distribution of a shock directly after a QRS complex is noticed. In patients with certain arrhythmia, directing a random counter shock transmits the danger of inducing ventricular fibrillation. In such cases, harmonized cardioversion is typically indicated. A defibrillation shock transported during the susceptible period of the cardiac cycle may cause ventricular fibrillation.

The Association for the Advancement of Medical Instrumentation (AAMI) endorses that defibrillators functioning in harmonized mode deliver the harmonized shock before 60 ms after the top of the R wave. Since the susceptible period directly follows this 60 ms “window”, distribution of the shock must not occur more than 60 ms after the peak of the R wave.

Appraisal: What is Non-invasive Pacing

 The defibrillator/monitor is chiefly used as a demand-mode pacer, which transports pacing treatment founded on the ECG signal assimilated through its ECG patient cable made by ECG Patient cable Manufacturers. Placing another machine in the ECG signal path may consequence in unforeseen distortions of the pacer-driven ECG waveform, which may, in turn, affect pacing treatment.

“Slaving” another device from the analog yield of a defibrillator/monitor

 When a defibrillator/monitor is used in combination with another machine, the medical staff may want to view the patient’s ECG signal on all components. A typical construction would permit the patient to be linked to the defibrillator/monitor plus have the aptitude to quickly intersect an ECG Patient cable to the input of another machine (s). This is the presently favored technique for synchronized Cardioversion or pacing.

“Slaving” The Defibrillator/monitor from the analog yield of another machine.

 It may be unreasonable or unwanted to place additional monitoring electrodes on the patient. This “slaving” is a suitable technique If you desire to also view the ECG waveform on the defibrillator/monitor, though, the distribution of Synchronized Cardioversion treatment or pacing with machines unified in this fashion can result in precise patient dangers and thus must not be done. To use the defibrillator/monitor during a cardioversion process or pacing, the ECG Patient cable must be linked straight to the defibrillator/monitor.

Non-invasive Pacing

 VITAL: The defibrillator/monitor must not have slaved from the analog yield of another machine when used for pacing. When pacing, the defibrillator/monitor must be linked straight to the patient through the ECG Patient cable supplied by the ECG Patient cable Suppliers.

Patient Risk Current

 When intersecting medical equipment, chassis seepage or danger currents may be augmented. To safeguard patient security, the hospital biomedical staff must perform electrical security testing on the complete system to ensure that safe current restriction are not surpassed.